- Flame without match
- Innovative safety Technologie
- Simple handling
- Cost-saving and environmentally friendly
Multi-Use
The Flame 100 is ideally suited for all flame related applications in the laboratory. The 15 millimeter precision flame allows graduated heating of ental tools as well as safe sterilization of microbiological instruments. Flame size and intensity can be adjusted infinitely.
Easy Operation
The safety Bunsen burner Flame 100 activates immediately with the push of a button. No match or a pilot flame are required. Optionally, the Flame 100 can also be operated by a foot pedal* or an external IR-Sensor*. The Flame 100 is suitable for stationary natural gas and propane/butane gas supplies as well as gas cartridges or gas cylinders. Numerous adapter systems are optionally available.
Reliable safety
The proven Safety Control System (SCS) is also incorporated in the Flame 100. All potential hazards are constantly monitored and, if necessary, protective measures – such as the shutting off the gas supply – are activated. In addition to the ignition and flame control function and the overheating protection the safety package also features the continuous burner head control BHC. BHC unfailingly identifies burner head clogging by liquids or solid substances. Compared with a conventional Bunsen burner, the Flame 100 with the safety standard „SCS“ eliminates the risk of a gas leak or an explosion.
Efficient
The Flame 100 is economical to operate and attractively priced for any laboratory, thus making savings for both your udget and the environment. Compared with traditional Bunsen burners, the gas consumption is reduced to a minimum, because the flame burns only when it is really needed!
Compact and Rugged
The housing and burner tube are made of stainless steel. The removable burner tube makes cleaning easy and the small foot print is an advantage for any lab bench.
Flame 100 - for a safe flame in the laboratory!
*optional
Technology | Microprocessor |
Programs Button: Foot pedal: | Start-Stop with timer, 60 min Standard (flame during pressed foot pedal, foot pedal optional)
|
Safety features Safety Control System (SCS) with gas safety cut off: | ignition and flame control, temperature monitor burner head clogging monitor (BHC) automatic unit switch off after 4h
|
Gas supply and consumption Gas connection: Gas types: Connected load: Continuous cartridge operation: | 1/4” left with gas filter II2ELL3B/P: natural gas E/LL,18 - 25 mbar liquid gas, 47,5 - 57,5 mbar 43 g/h liquid gas, 49 l/h natural gas (LL) CV 360 - 65 min CG 1750 - 210 min C 206 - 230 min CP 250 - 305 min CV 470 - 550 min CV 270 - 280 min
|
Temperatures Flame temperature: Temperature threshold level: | 1200 °C on liquid gas 1170 °C on natural gas (E) 0.55kW liquid gas, 0.50kW natural gas (E), 0.40kW natural gas (LL)
|
Electrical Power consumption: Power connection: | 2 VA (stand by max. 0.1 VA) 100-240V / 50/60Hz / max. 0.3 A; 9V DC / 1.3A Level 6 (stand by power consumption max. 0.1W)
|
Mechanical Burner tube: Measurements (w x h x d): Height with burner tube: Weight: | removable, stainless steel 89 x 34 x 88 mm 94 mm 385 g
|
Licenses DIN-DVGW Reg.-No.: CE: EU guidelines: | NG-2211AS0167 EN 61326-1, EN61010-1, EN61010-2-010 2014/30/EU, 2014/35/EU, 2011/65/EU
|
The range including tools for a simple start-up of the Flame 100.
- Safety laboratory Bunsen burner Flame 100 with button function
SCS (Safety Control System) with BHC (Burner Head Control)
Removable burner tube
Holding device for 2 inoculation loop holders
Instruction manual and 2-year warranty - Tubing connector with swivel nut
- Switching power supply for global usage 100 - 240 V / 50/60 Hz (level 6)
- Screwdriver for electrode holder
- Wrench 17 mm for gas connection
- Nozzle natural gas (installed in the Flame 100)
- Nozzle for propane / butane gas
Gas-Safety Adapters, Tubes, Connectors, Cartridges