SteriMax smart

NEW ! Sterilizing by IR-HotSpot !

SteriMax smart - Infrared Loop Sterilizer. High-Speed Annealing and Sterilizing without flame.




Item Number: 4.001.000 / 4.002.000 / 4.003.000

• Instantly ready to work without any unnecessary warm-up
• Auto-Start through touchless IR-Sensor Technology
• Adjustable sterilization and cool-down timers for two users by touch operation
• Annealing tube made of special quartz glass
• Very simple cleaning of the device and the annealing tube
• Efficient use of energy and minimum heat Dissipation thanks to the Thermocontol Technology
• Cool-Touch housing made of stainless steel and safety glass
• Tested and certified for photobiological safety of lamps and lamp systems according to EN 62471 by BG ETEM - Risk-free for skin and eyes

High-Speed Annealing and Sterilizing the Smart Way
Are you looking for a fast, efficient and flameless product to anneal or sterilize your inoculation loops and delicate micro instruments? Then our electrical SteriMax smart should be your first choice – ideally suited for use in any type of lab, anaerobic work environment and Laminar Flow Cabinet.

Smart Infrared Light – Instantly Ready to Work
Specifically focused infrared light generates an IR-HotSpot in which your inoculation loop is sterilized at temperatures from 750 to 1000°C after only 5 to 10 seconds. The SteriMax smart just needs to be plugged in and will be immediately ready to use, without any unnecessary warm-up period. Our highly sophisticated annealing tube is based on many years of experience. It is made of special quartz glass. The tube is closed at one end preventing any risk of  ontamination with pathogenes. The annealing tube can be slid out in a matter of seconds for easy cleaning.

Smart Operation – Touchless Working
SteriMax smart allows you to fully focus on your core task. Hardly any attention is needed for inserting your inoculation loop into the annealing tube – which is easily accessible from the front of the flat glass panel. Touchless IR-Sensor technology starts each sterilization process automatically. Adjustable timers signal both the completion of sterilization and cooling via the display and an audible signal. After that your inoculation loop is ready for use. Both timers can be adjusted and recalled individually and up to the second for two users by gently touching the panel.

Smart Use of Power – Minimum Heat Dissipation
’Thermocontrol’ is the key element for an efficient use of power. It allows for quick but smooth softstart of the infrared light, a short heating period and the re-use of residual heat for the next sterilization process, which will be respectively shorter. Whenever you pause, your energy consumption and heat dissipation will pause as well.

Smart Design – Outstanding Features
SteriMax smart comes in an elegant stainless steel housing with a touch safety glass control panel designed for extreme laboratory environments with many unique features.


Operation
IR-Sensor:
Touch operation:


sensor range 8-stage adjustable
ON / OFF, selection for 2 users
sterilization timer 5 - 10 s
inoculation loop cooling timer 0 - 25 s with
audible signal (disengageable)

Thermocontrol:
Status Display:
Sterilization at start-up:

dynamic sterilization timer
optical display of the sterilization progress
sterilization of the annealing tube before usage (disengageable)

Annealing tube:
Sterilization temperature:

Quartz glass
750 °C - 1000 °C (1382 °F - 2012 °F)

Safety features
Residual heat display:
Automatic unit switch off:
Temperature monitor:


indicates a hot annealing tube
selectable 15 min., 30 min., 1 h
thermal circuit breaker

Electrical
Power supply:


Power consumption:
Protection class:
Heat source:


230 V mains voltage: 220 - 240 V  50/60 Hz  2.6 A
120 V mains voltage: 112 - 128 V  50/60 Hz  5.1 A
100 V mains voltage: 95 - 105 V  50/60 Hz   6 A
approx. 600 W / short time (5 - 10 s)
IP 20
IR halogen spot light

Mechanical
Quartz glass annealing tube:
Casing:
Measurements (w x h x d):
Weight:


outer-Ø 19 mm, inner-Ø 16 mm, length 112 mm
heat resistant glass / stainless steel, UV and solvent resistant
110 x 170 x 180 mm
approx. 1800 g

Licenses
CE:
EU guidelines:


EN 61326-1, EN 61010-1, EN 61010-2-010, EN 62471
2014/30/EU, 2014/35/EU, 2011/65/EU, 2006/25/EG

Information on radiation
Emission limit:


The emission limit „risk free“ is respected.
The device is assigned to the risk free group
according to EN 62471.
Tested for photobiological safety of lamps and lamp systems by BG Energie, Textil, Elektro, Medienerzeugnisse - Risk-free for skin and eyes

Protected Design

DM/082 011


The range for a simple start-up of the SteriMax smart.

  1. SteriMax smart
    Infrared light technology
    1Quartz glass annealing tube installed in the unit
    Holding device for 3 WLD-TEC inoculation loop holders
    Operating manual
    2-year warranty (except parts to wear and tear)
  2. Power supply
  3. 1 inoculation loop Ø 3 mm
  4. 1 inoculation loop holder
  5. 1 Replacement - Quartz glass annealing tube

    for 230 V mains voltage
    for 120 V mains voltage
    for 100 V mains voltage

    Art.-No. 4.001.000
    Art.-No. 4.002.000
    Art.-No. 4.003.000

     


Trays
Miscellaneous

Durch die weitere Nutzung der Seite stimmst du der Verwendung von Cookies zu. Weitere Informationen